SPENCER 170NXT LUNG VENTILATOR EV01170NB

Electronic lung ventilator for non-invasive ventilation of adult and pediatric patients.

Spencer 170 NXT is an electronic emergency ventilator for non-invasive ventilation of adult and pediatric patients. It is a cost-effective and compact electronic ventilator with a strong protective shell, a lever button switch, and two alarm levels for low battery.

Specific Features:

  • Protection against accidental activation of the AIR MIX and ON/OFF selection levers
  • Color-coded manometer for an easy identification of pressure ranges.
  • The positioning of input and output ports on the side of the device, makes connecting and disconnecting easier and offers greater protection to the elements connected to the device
  • Control knobs on high contrast background for a better identification of the set values
  • The aluminum structure is compact and allows extreme reduction of weights
  • The placing and the shape of adjustment knobs make the setting of parameters easier in all operative conditions
  • Constant I/E ratio
  • Alarms for ventilation monitoring and device management
  • The presence of one ventilation mode makes it simple and easy to operate.

Reference Standards:
EN 794-3
EN 1789
EN 60601-1
EN 60601-1-2
EN 60601-1-8
UN/ECE Regulation 10 – Approval number E24 10R-041994

Standard Equipment:

  • Patient circuit
  • Face mask
  • O2 Connection tube

Optional accessories:

  • AC/DC adapter 100/240 VAC 50-60Hz.

Technical data:

  • Materials: Al, Steel, Brass, PE
  • Dimensions: 270x177x210 ± 20 mm
  • Weight: 3,44 ± 0,2 kg
  • Alarm’s: High pressure – Pneumatic supply – Low battery (Two stages)
  • Alarms sound press. Level: 80 ± 10db
  • Audio mute time: 100 s
  • Frequency: From 5 to 40 bpm
  • Flow: From 2 to 20 L/min
  • Overpressure valve adjustment: From 20 to 60 cmH2O
  • I/E Ratio: 1:2
  • O2 Concentration: 60% or 100%
  • Manometer: From -20 to +80 mbar
  • Pneumatic connection standard: UNI
  • IP protection: IP24
  • Classification: Class II or internally powered
  • Power supply: External by 12V cigarette lighter or approved AC/DC adapter 100/240VAC 50-60Hz (optional).
  • Battery operating time: About 3 hours
  • Applied part: Type BF
  • Pneumatic supply: From 280 to 600 kPa
  • Non-rebreathing valve: 60cmH2O

Class IIb Med.Device according to Dir. 93/42/CEE
SPENCER 170 NXT ELECTR. PULM. VENT. ENG PAN.
EV01170/NB